Specifications; Symbol Identification - Dentsply Sirona Cavitron Select SPS Instructions For Use Manual

Table of Contents

Advertisement

14: Specifications

Cavitron® Select™:
100-240 VAC Systems
Electrical Voltage
30 VDC +/- 5%
Current
2 amps DC Max
Frequency
50/60 Hertz
Pressure
Water Pressure
20 – 40 psig
Water Flow Rate
Minimum setting (CCW)
0 – 15 ml/min
Maximum setting (CW)
>45 ml/min
Weight
425 g / 15 oz.
Dimensions
Height: 2.4 in.
Width: 3-1/8 in.
Depth: 7 1/2 in.
Handpiece Cable: 72 in.
Foot Control Cable: 72 in.
Power Cord: 2 m
Water Supply Line: 96 in.
Footswitch
Not for operating theatres
Protection Class IPX1
Operating
Temperature: 10 to 40 Deg. Celsius
Environment
Relative Humidity: 30% to 75% (non-condensing)
Transport and
Temperature: -40 to 70 Deg. Celsius
Storage Conditions
Relative Humidity: 10% to 95% (non-condensing)
Atmospheric Pressure: 500 to 1060 hPa
Cavitron® Select™ Power
Cavitron® Select™ Reservoir
Supply (674022001)
Pump:
100 – 240 VAC
30 VDC +/- 5%
1,6 amps Max @
Not applicable
100 – 240 VAC
47 – 63 Hertz
Not applicable
Pump Pressure
Not applicable
15 – 19 psig
Not applicable
40 ml/min
550 g / 19.4 oz. Max
513 g / 18 oz.
Height: 14,3 cm / 5.7 in.
Height: 10,2 cm / 4.0 in.
Width: 7,6 cm / 3.0 in.
Width: 8,4 cm / 3.3 in.
Depth: 14,3 cm / 5.7 in.
Length: 30 cm / 11.8 in.
Power Cord: 213 cm / 84
in.

15. Symbol Identification

TYPE B APPLIED PART EQUIPMENT
Footswitch not for operating theatres
IPX1
Protection Class- IPX1
IPX1 Classification of ingress of water
Consult Instructions for Use
Consult
Instructions
for Use
Footswitch
This symbol is a mandatory marking for devices entering the European market to indicate conformity with the
essential health and safety requirements set out in European Directives. The symbol may be accompanied by a four-
digit identification number of the notified body.
Dispose of in accordance with the Waste Electrical and Electronic Equipment Directive 2002/96/EC of the European
Parliament and the Council of the European Union
Direct Current
Caution, Consult Accompanying Documents
Caution
Underwriters Laboratories Compliant
Indicates 60601-1-2 EMC Compliant
VDE Mark
English - 18