UNI-T UDP3000S Series Programming Manual page 18

Programming dc power supply

Advertisement

Programming Manual
For Example::OUTPut:OVP:STATe? CH1
Return:ON
:OUTPut:OCP:VALue
Format::OUTPut:OCP:VALue [CH1|CH2|CH3|SER|PARA,] {<curr>|MINimum|MAXimum}
Function:Set the overcurrent value of a certain channel and set the channel as the current channel.
Description:If the channel parameter is ignored, then the current channel which is the channel need to be
controlled by default. "CH1, CH2"can only be selected in normal mode; "SER" can only be selected in serial
mode; "PARA"can only be selected in parallel mode.
For Example::OUTPut:OCP:VALue CH1, 5.1
:OUTPut:OCP:VALue?
Format::OUTPut:OCP:VALue? [CH1|CH2|CH3|SER|PARA]
Function:Query the overcurrent value of a certain channel.
Description:If the channel parameter is ignored, then the current channel which is the channel need to be
queried by default.
For Example::OUTPut:OCP:VALue? CH1
Return:5.100
:OUTPut:OCP[:STATe]
Format::OUTPut:OCP[:STATe] [CH1|CH2|CH3|SER|PARA,] {0|1|OFF|ON}
Function:Set the overcurrent protective switch of a certain channel and set the channel as the current channel.
Description:If the channel parameter is ignored, then the current channel which is the channel need to be
controlled by default. "CH1, CH2"can only be selected in normal mode; "SER" can only be selected in serial
mode; "PARA"can only be selected in parallel mode.
For Example::OUTPut:OCP:STATe CH1, ON
:OUTPut:OCP[:STATe]?
Format::OUTPut:OCP[:STATe]? [CH1|CH2|CH3|SER|PARA]
Function:Query the overcurrent protective switch of a certain channel.
Description:If the channel parameter is ignored, then the current channel which is the channel need to be
queried by default. It returns"ON | OFF".
For Example::OUTPut:OCP:STATe? CH1
Return:ON
6. MEASure
:MEASure:ALL[:DC]?
Format::MEASure:ALL[:DC]? [CH1|CH2|CH3|SER|PARA]
Instruments.uni-trend.com
UDP3000S Series
18 / 49