Download Print this page

UNI-T UDP3000S Series Programming Manual page 17

Programming dc power supply

Advertisement

Programming Manual
For Example::OUTPut:STATe? CH1
Return:ON
:OUTPut:CVCC?
Format::OUTPut:CVCC? [CH1|CH2|CH3|SER|PARA]
Function:Query the status of the constant voltage and constant current of a certain channel.
Description:If the channel parameter is ignored, then the current channel which is the channel need to be
queried by default. It returns"CV | CC".
For Example::OUTPut:CVCC? CH1
Return:CV
:OUTPut:OVP:VALue
Format::OUTPut:OVP:VALue [CH1|CH2|CH3|SER|PARA,] {<vol>|MINimum|MAXimum}
Function:Set the overvoltage protective value of a certain channel and set the channel as the current channel.
Description:If the channel parameter is ignored, then the current channel which is the channel need to be
controlled by default. "CH1, CH2"can only be selected in normal mode; "SER" can only be selected in serial
mode; "PARA"can only be selected in parallel mode.
For Example::OUTPut:OVP:VALue CH1, 5
:OUTPut:OVP:VALue?
Format::OUTPut:OVP:VALue? [CH1|CH2|CH3|SER|PARA]
Function:Query the overvoltage protective value of a certain channel.
Description:If the channel parameter is ignored, then the current channel which is the channel need to be
queried by default.
For Example::OUTPut:OVP:VALue? CH1
Return:5.00
:OUTPut:OVP[:STATe]
Format::OUTPut:OVP[:STATe] [CH1|CH2|CH3|SER|PARA,] {0|1|OFF|ON}
Function:Set the overvoltage protective switch of a certain channel and set the channel as the current channel.
Description:If the channel parameter is ignored, then the current channel which is the channel need to be
controlled by default. "CH1, CH2"can only be selected in normal mode; "SER" can only be selected in serial
mode; "PARA"can only be selected in parallel mode.
For Example::OUTPut:OVP:STATe CH1, ON
:OUTPut:OVP[:STATe]?
Format::OUTPut:OVP[:STATe]? [CH1|CH2|CH3|SER|PARA]
Function:Query the overvoltage protective switch of a certain channel.
Description:If the channel parameter is ignored, then the current channel which is the channel need to be
queried by default. It return"ON | OFF".
Instruments.uni-trend.com
UDP3000S Series
17 / 49

Advertisement

loading

This manual is also suitable for:

Udp3305s