Open The Measuring Instrument - TESTO 300 Pro Smoke Edition Manual

Combustion analyzer
Table of Contents

Advertisement

Find Quality Products Online at:

10.6 Open the measuring instrument

Only open the measuring instrument when this is required for maintenance
purposes (replacing gas sensors).
The measuring instrument must not be connected to a mains socket via
the mains unit. The measuring instrument must be turned off.
When opening/assembling the instrument, take care not to lose any
screws that have been removed. It is advisable to place a cloth on the
work surface.
1
Place the instrument on its front so that the back is facing upwards.
2
Using a torx screwdriver (size T 10),
remove both housing screws on the
top of the instrument.
The instrument may be damaged by incorrect removal of the housing
screws!
- Only remove the two housing screws on the top of the instrument. The
other four screws must be left as they are.
3
Unlock the operating module in the
direction of the arrow.
4
Remove the operating module.
GlobalTestSupply
www.
10 Maintenance
ATTENTION
.com
65
sales@GlobalTestSupply.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

300 pro300 smoke edition300Pro

Table of Contents