10 Maintenance
Weak mix of acids in the condensate!
Minor injuries!
- Avoid skin contact.
- Make sure that the condensate does not run over the housing.
Condensate entering the gas path!
Damage to sensors and flue gas pump!
- Do not empty condensate trap while the flue gas pump is in operation.
1
Open the condensate outlet on the
condensate trap.
2
Let the condensate drain out.
3
- Wipe off any drops still on the condensate outlet with a cloth and close
the condensate outlet.
The condensate outlet must be completely closed, otherwise incorrect
measurements could occur due to infiltration of external air.
64