14. The LED Display now shows the Temperature Display Impedance mode as Enabled
or Disabled ("AC0" = Disabled and "AC1" = Enabled). Choose one of the following:
a) Press and release the Program Key (
(Disabled or Enabled).
b) Press and release the Scroll Up (
or Enabled). Press and release the Program Key to proceed to the next step.
Exiting Set-Up Mode
15. The LED Display now reads "End". The Set-Up Mode procedure is now complete.
Choose one of the following steps:
a) Press and release the Scroll Up (
and return to normal operation.
b) Press and release the Scroll Down (
of the Set-Up Mode procedure. Go back to step 4.
Factory Settings
The ST 45 and ST 55 systems come equipped with a number of features, which may be adjusted as
desired by the user. Listed below are the features and factory settings of each. To change and/or
learn about any of these features, refer to the applicable part of the "Customizing Your System"
section of this manual.
Feature
Password
Default Temperature Scale (°C/°F)
"HI" (Upper) Temperature Limit
"LO" (Lower)Temperature Limit
Set Temperature
Tip Offset Constant
Temperature Setback
Auto Off
Temperature Display Impedance Mode
Corrective Maintenance
System Accuracy and Calibration
No calibration adjustments are necessary to maintain the accuracy of the system.
LED Display Message Codes
Listed below are message codes, which, may be shown on the LED Display if a mistake is made by
the operator (e.g., wrong Password entry) or if the system has malfunctioned.
LED Display Message
©2001 PACE Inc., Laurel, Maryland
All Rights Reserved
Factory Setting
None Entered
°F for 115 VAC Systems
°C for 230 VAC Systems
454 °C (850 °F)
204 °C (400 °F)
"OFF"
"0"
Enabled, 30 minutes
Enabled, 30 minutes
Enabled
Table 1: Factory Settings
The incorrect password has been entered. The displayed
message will time out after 6 seconds and revert to normal
operation. Enter the correct password.
No handpiece is connected to the power receptacle. Connect
handpiece.
) to keep the currently stored setting
) Key to change the stored setting (Disabled
) Key to exit Set-Up Mode
) Key to return to the start
Description
Page 16 of 20