Parameter
LAN Page 1/5 Network configuration
Mode
IP Address
Subnet mask
DNS servers
Gateway
Time server
LAN Page 2/5 Network folder
Network share
User name
Password
Input path
Timer
Output path 1
Output format 1
Output path 2
* indicates default values
Description
Selection of configuration mode
"Manual" configuration mode: entry of
network parameters
"DHCP" configuration mode: display of data
obtained from the DHCP server
Input of a network share following the syntax
\\<Fileserver>\<Folder>
The file server can be entered via its IP address
or its network designation.
Input of a user name to get access to the file
server.
Depending on the configuration of the
network, it may be necessary to specify the
network instance that manages/grants access
authorization in addition to the user name. In
this case use the syntax <user
name>,domain=<domain>. Here, <domain>
stands for the IP address or the network name
of the domain controller.
Input of a user name and the corresponding
password to get access to the file server.
Path specification for an input folder (for
example, for software update via the network)
Definition of a time constant for the periodic
reading from the input folder
Definition of an output path 1 for storage of
measurement data on a network folder. Please
note, that ZEISS VISUREF 150 does not create
any folders. The folders should already have
been created in the network.
Definition of an output path format for storage
on output path 1 of the network folder
Definition of output path 2 for storage of
measurement data in a network folder.
Please note that ZEISS VISUREF 150 does not
create any folders. The folders should already
have been created in the network.
Operation of the device
Selection and input options
Manual
DHCP*
Text input (in "Manual"
configuration mode)
Text input
Example: \\192.168.99.10\RDx or
\\dfszorg445\RDx
Text input
Example:
guest1, domain=server1
Text input
Default: device\input*
Selecting the timer
Default: 5s*
Text input
Default: data\txt*
v1.6*
v1.7
XML JOIA
PDF simple
PDF enhanced
Other formats depend on
installation
Text input of output path
Default*: data\xml
000000-2227-967-GA-GB-281021