Electrical connections meas. cell
Gas outlet measuring cell
Gas inlet measuring cell
Fastening screw for measuring cell
Figure 9
Unsolder the brown and yellow cable from the terminals 1 (see figure 9) on the back side of the measuring cell; do
not overheat the terminals; mark the cables accordingly;
Disconnect the tubing for the sample gas outlet 2 and elbow fitting at the sample gas inlet;
Loosen the fastening screw of the measuring cell 4 with a screw driver;
Push cell carefully out of the magnetic field with both thumbs. (Attention! The necessary power to this is not
negligible.)
Remove cell;
Exchange only with measuring cells of the same type;
The mounting of the measuring cell is to be effected in reverse order; take care of the correct dumpbell position!
W A R N I N G !
32
Transmitter unit
When using PTFE ferrules, they always have to be renewed.
PMA30 | 1.01.03
Heater element
Transmitter board
Temperature cut out at 72 °C
Temperature sensor