Figure 9 Transmitter Unit - M&C PMA Series Instruction Manual

Oxygen – analyzer
Hide thumbs Also See for PMA Series:
Table of Contents

Advertisement

Electrical connections meas. cell
Gas outlet measuring cell
Gas inlet measuring cell
Fastening screw for measuring cell
Figure 9
 Unsolder the brown and yellow cable from the terminals 1 (see figure 9) on the back side of the measuring cell; do
not overheat the terminals; mark the cables accordingly;
 Disconnect the tubing for the sample gas outlet 2 and elbow fitting at the sample gas inlet;
 Loosen the fastening screw of the measuring cell 4 with a screw driver;
 Push cell carefully out of the magnetic field with both thumbs. (Attention! The necessary power to this is not
negligible.)
 Remove cell;
 Exchange only with measuring cells of the same type;
The mounting of the measuring cell is to be effected in reverse order; take care of the correct dumpbell position!
W A R N I N G !
32
Transmitter unit
When using PTFE ferrules, they always have to be renewed.
PMA30 | 1.01.03
Heater element
Transmitter board
Temperature cut out at 72 °C
Temperature sensor
www.mc-techgroup.com