In case there are minimally different positions of the dumpbells inside the measuring cells when mounting a
new cell, it is absolutely necessary to adjust the zero point mechanically. The PMA30 housing can remain open
for this purpose;
Fix earthing cable (geen-yellow) of the transmitter unit;
Put on 18-pole flat ribbon cable to the transmitter;
Put on the 2-, 3- and 4-pole green plug connections on the main board;
Put black isolation cover over the transmitter unit.
20.2
MECHANICAL ZERO POINT ADJUSTMENT
The mechanical zero point adjustment has to be done as described in the following.
Before switching on the analyzer, set range selector to 30 %. Control zero point of the analogue display and adjust
it if necessary to 0 % via the adjustment screw below the digital display (Attention! dependent of position: observe
position of use);
Switch on the analyzer via external switch. Normally, the analogue display will indicate a value of 21 % oxygen
because the cell is filled with ambient air;
The warming up is indicated by the permanently beaming LED in the front plate of the analyzer; after approx. 30
minutes, the flashing display signalizes that the required operating temperature has been reached;
Set the potentiometer for zero point and span value in the mid position; you can do this by turning the
potentiometer with a screw driver fully to the left and then turn it five turns back to the right;
Charge the analyzer with a zero gas volume flow of approx. 40 l/h;
With a precise adjustment of the zero point, the analogue display should indicate 0.0 vol% oxygen;
If this does not happen, please execute the following steps: