Oxygen analyzer Series PMA
Materials in contact with sample gas
Sample gas connection
Flow alarm
Status alarm
Protection / electrical standard
Housing / colour
Dimension / weight
* Please specify with order.
**The minimized error of < 0.1 vol% for the reproducibility deviation can only be selected if the measurement and calibration
ranges are identical. If the instrument is switched to another measuring range, a value of < 1 % of the measuring range is to
be expected.
8.1

OPTIONS

Options
Extra charge for PMA 30 as solvent resistant execution with measuring cell type PMC-1LB. Material: O-
03A9165
rings out of Kalrez.®
02A9005
Extra charge for PMA 20/30 in chlorine resistant execution with a special measuring cell type PMC-
1CL2. All gas bearing tubes PTFE/PVDF equipped with purge gas connectors.
Extra charge for electrical zero suppressing with process pressure compensation type SD/PD for
03A9535
PMA30, pressure range 0.6 bar, media wetted parts: stainless steel, PVDF, FKM. Not with SIL-
certification!
Extra charge for process pressure compensation type PD for PMA30, pressure range 0.6 bar, media
03A9530
wetted parts: stainless steel, PVDF, FKM. Not with SIL-certification!
03A9150
Wall mount housing out of aluminum type 30-19 with 19"-insertion system (3U/42TE), incl. clamp
box, clamp boards and filter, with window. protection class: IP65.
01A9165
Extra charge for changed measuring range to 2.5/5/10/25/100 vol% O
italmeter (only for PMA30/D).
10
Version PMA 30/D, PMA 30A/D
Platinum, Glass, PTFE, PVDF, Stainless Steel 316Ti, FKM, Epoxy resin
1/8" NPT i
thermo-conductive flow sensor downstream mounted after measuring cell
2 x change over contact, potential free, max. 48 V DC, 0.5 A (min. switching ca-
pacity 50 mW) for flow min-max.-fault, transmitter temperature < 45 °C >60 °C,
light source defective, coupling state, measuring range exceeding, range se-
lection switch defective, power supply error, mains voltage breakdown (12
times internal LED-indication)
IP 20 EN 60529 / EN 61010
3U/28HP - 19-inch cassette European standard/silver
Height 3U, width 28HP, length 230 mm + approx. 60 mm installation space /
approx. 2.5 kg
PMA30 | 1.01.03
, for recorder outlet and dig-
2