13.3.6

DESCRIPTION 0-10 V OUTLET

0-10 V always correspond to 100 % O
For a load of > 100 KOhm
Linear from -0.5 V to 11 V
General evaluation of the 0-10 V signal:
Measuring Range
Lower deviation
The connection happens at the 9-pole D-sub-connector X4. (see figure 6 and 7)
13.3.7

STATUS SIGNAL

The status signal contains a collective fault message:
Power supply error, voltage breakdown
Flow min. fault
Flow max. fault (can be switched off: JP3)
Light source short circuit/line breakage
Selection switch defective
Transmitter temperature < 45 °C/> 60 °C (can be switched off: JP8/JP7, 2+3 on, 1+2 off)
Coupling state (can be switched off: JP6, 2+3 off, 1+2 on)
Fault High/Fault Low (connectible: JP14/JP12). Measuring range lower deviation and exceedance
more than +3.125 % resp. -4.375 %
An allocation of the existing failures is possible via 12 internal LED's on the main board (see figure 12). The board
becomes visible after removing the cover plate. To do this remove the two screws in the lid.
The status signal is accessible via a potential free switching contact (status contact) as well at X1 (D-sub 25-
pole) as also at X4 (D-sub 9-pole) (see figure 6 and 7).
Figure 7
20
2
Measuring Range
(corresponds to > 20.5 mA resp. < 3.2 mA)
,
Pin assignment of D-sub-connector X4
PMA30 | 1.01.03
independent of the measuring range
Measuring Range
exceedance