NOTE
Overflowing bath fluid due to externally connected systems!
If the externally connected system is higher than the temperature control
system, bath fluid can flow back and overflow when switched off.
Position the connected external system on the same level or lower than
the temperature control system
Position a shut-off valve or Solenoid valve between the external system
and the temperature control system as backflow protection.
NOTE
Damaged hoses due to kinking!
Hoses are damaged by kinking and can may leak.
Lay hoses with large radii
Avoid kinking of the hoses

7.3.1 Connect an external system with screw connections

This section describes how to connect an external, closed system to the device
using screw connectors.
 Open-end wrench, 17 mm
 Open-end wrench, 19 mm
 Torque wrench
 The circulator is equipped with the optional assembly frame or pump set.
 The temperature control hoses of the external system are equipped with
M16x1 female and barrel gaskets.
1. Remove the union nuts on the pump connections.