Adjust External Temperature Sensor - Julabo 9021713.S1.02 Original Operating Manual

Heating immersion circulators, heating circulators with open baths, refrigerated circulators
Table of Contents

Advertisement

Operation

8.11.2 Adjust external temperature sensor

74
Find Quality Products Online at:
3. From the <Install unit> submenu, select the menu option
<Adjust sensor>.
4. Set the internal mode to inactive.
 The input fields are unlocked.
5. Enter the desired setpoint for internal in the first line and confirm your
entry.
6. Enable the number field highlighted in blue on the first line.
 The temperature control application starts. Let the reference thermometer
settle for 10 minutes.
7. Enter the measured reference value in the first line in the Cal. column.
8. If necessary, repeat steps 5 to 7 for a 2-point or 3-point adjustment.
9. Activate the mode for the desired number of measuring points and press the
[Save] softkey.
 The stored correction values are active immediately.
 The internal temperature sensor is adjusted.
Measurable temperature differences can also occur in a connected application.
As with the internal temperature sensor, adjustment of the external temperature
sensor can improve accuracy of the temperature control application.
 The unit is switched on.
 External control is active.
 An external temperature sensor is connected.
 The bath tank is at least two thirds full.
 A calibrated reference thermometer is available.
1. Hang the reference thermometer in the connected external bath tank.
2. Bring up the <Main menu>.
3. From the <Install unit> submenu, select the menu option
<Adjust sensor>.
4. Set the mode for external to inactive.
 The input fields are unlocked.
5. For external, enter the desired setpoint in the first line and confirm your
entry.
6. Enable the number field highlighted in blue on the first line.
 The temperature control application starts. Let the reference thermometer
settle for 10 minutes.
7. Enter the measured reference value in the first line in the Cal. column.
8. If necessary, repeat steps 5 to 7 for a 2-point or 3-point adjustment.
9. Activate the mode for the desired number of measuring points and press the
[Save] softkey.
 The stored correction values are active immediately.
 The external temperature sensor is adjusted.