Clean Device - Julabo 9021713.S1.02 Original Operating Manual

Heating immersion circulators, heating circulators with open baths, refrigerated circulators
Table of Contents

Advertisement

Maintenance
9.6
78
Find Quality Products Online at:

Clean device

The circulator and bath tank, and also a cooling machine if connected, should
be cleaned from time to time.
In addition to this, the device must be appropriately decontaminated if
hazardous substances have been spilled on or into the device.
 Lint-free cloth
 Mild cleaning agent
NOTE
Damage to the electronics due to water penetration!
Ingress of water can damage electronic components of the device and thus
lead to failure of the device.
Clean the outside of the device with a damp cloth only
Prevent water from entering the device
 The device is switched off and disconnected from the mains voltage.
1. Allow the device to cool down to room temperature.
2. Completely drain the bath fluid.
3. Clean the surface of the bath tank and the circulator with a damp cloth.
 Some dish detergent may also be used for cleaning. If in doubt, ask
technical service for alternative cleaning mediums.
4. Take the venting grid off the cooling machine.
5. Carefully vacuum the dirt off the condenser.
 Make sure that the fins of the condenser are not damaged.
6. Place the venting grid back on.
 The device has now been cleaned.