Reg/Eprog Socket - Julabo 9021713.S1.02 Original Operating Manual

Heating immersion circulators, heating circulators with open baths, refrigerated circulators
Table of Contents

Advertisement

Find Quality Products Online at:
In the <Analog module> menu, a defined state can be selected as the trigger
for the <Alarm output>.
When the respective state is reached, pins 2 and 3 are connected:
Standby
Alarm
Alarm + Standby
When the respective state is reached, pins 1 and 2 are connected:
Standby/Inverse
Alarm/Inverse
Alarm + Standby/Inverse
Alarm output technical data
Maximum switching capacity
Maximum switching voltage
Maximum switching current

5.5.2 REG/EPROG socket

The REG/EPROG socket has three signal outputs and one signal input.
A signal transmitter connected to the programmer input can specify the setpoint
temperature or performance as a variable. The input signal can be a voltage
source or a current source.
The signal outputs can be used at the same time. Two of the three output
channels are designed as voltage outputs, the third as a current output.
REG/EPROG socket diagram
GlobalTestSupply
www.
30 W/25 VA
30 VDC/25 VAC
1 A
[1] Voltage output channel 1
(0-10 VDC, max. 10 mA)
[2] Voltage output channel 2
(0-10 VDC, max. 10 mA)
[3] Gnd for outputs (0 V)
[4] EPROG programmer
(0-10 VDC/0-20 mA, max. 24 VDC)
[5] Current output channel 3
(4-20 mA/0-20 mA, max. 24 VDC)
[6] Gnd for EPROG programmer (0 V)