Set Chiller Mode - Julabo 9021713.S1.02 Original Operating Manual

Heating immersion circulators, heating circulators with open baths, refrigerated circulators
Table of Contents

Advertisement

Find Quality Products Online at:
 The device is switched off.
 Bath fluid is filled.
1. Set the lever to a low level for internal circulation (e.g. position 2 from the
right stop).
2. Switch on the device and check whether the flow of the bath fluid suits your
application.
3. If necessary, adjust the flow until it suits your application.
 The flow of the bath fluid is set.
7.9

Set chiller mode

For a refrigerated circulator, the chiller mode is preset ex works to automatic
operation. Three different chiller modes can be selected in the operating menu:
Automatic mode: The controller calculates the duration of the heating
process in the event of a setpoint change of more than +5°K. Based on the
calculated heating duration, the control decides whether the refrigeration
unit is to be switched off or not, depending on demand.
Always on: The refrigeration unit is permanently switched on during
operation.
Always off: The refrigeration unit remains switched off during operation.
 The device is switched on.
 The cooling machine is connected.
1. Call up the <Main menu>.
2. Call up the submenu <Install unit> then the menu option
<Chiller mode>.
 The menu option can only be called up if the thermostat detects a
connected refrigeration unit.
3. Select the desired chiller mode.
 The chiller mode is set.