Status Commands; Status Messages - Julabo 9021713.S1.02 Original Operating Manual

Heating immersion circulators, heating circulators with open baths, refrigerated circulators
Table of Contents

Advertisement

Find Quality Products Online at:
Auxiliary
Parameter
parameters
out_hil_00
-xxx
out_hil_01
xxx

13.1.3 Status commands

Status commands are used to query the current status of the device.
Status commands
System response
version
Current firmware version
status
Return of status, error, warning, alarm

13.1.4 Status messages

Possible status messages from the device for a status query.
Status message
Explanation
00 MANUAL STOP
Device in standby mode manual operation
01 MANUAL START
Device in manual mode
02 REMOTE STOP
Device in standby mode, remote control operation
03 REMOTE START
Device in remote control mode
-08 INVALID
Device did not recognize the last received command
COMMAND
-09 COMMAND NOT
Last received command is not permissible in the operating mode
ALLOWED IN
CURRENT OPERATING
MODE
-10 VALUE TOO
The last set value is too small
SMALL
-11 VALUE TOO
The last set value is too large
LARGE
-13 VALUE EXCEEDS
The value is not within the set temperature limits
TEMPERATURE
LIMITS
GlobalTestSupply
www.
Setting
Setting of the actuating variable limit for cooling
capacity (0-100%)
Setting of the actuating variable limit for heating
capacity (0-100%)
.com
Appendix
89