Install the device at the operating location

This section describes how the device is set up at the installation location.
 The device has been transported to the operation location.
 The size and infrastructure of the operation location are suitable for device
operation.
1. If possible, position the device under an extraction system.
 Depending on the bath fluid, gases may be created at high temperatures.
 Recommended minimum distance of 1 m to other devices, to prevent
electromagnetic interference.
2. Place the device on a level, smooth, non-flammable surface.
3. Ensure that the device is securely positioned.
4. For refrigerated circulators: Ensure an open space in front of and behind the
device of at least 20 cm.
 The device is set up at the operation location.