7.1

Installation of the circulator

Danger of scalding due to leaks from the baths
The JULABO plastic baths are not solvent-resistant. JUALBO plastic
bath tanks are for water at a working temperature range from +20°C
to +100°C.
Do not contaminate the bath fluid with solvents.
Things to keep in mind during the installation process:
Risk of tipping due to improper transportation
Crushing, damage to the unit
Pos: 1 4 /Technis ch
8

Preparations for operating the device

8.1

Securing the immersion circulator

Danger of electric shock.
Carefully secure the immersion circulator on the bath vessel. Poorly
installed circulators can fall into the bath tank.
Have the unit checked by a service technician prior to re-use.
The heater must not be in contact with the wall or the bottom of the
bath tank. Minimum distance 15 mm.
Pull the plug to disconnect the unit from the power supply. Only then
take the immersion circulator out of the bath tank.
16
www.calcert.com
15
10
20
5
25
0
30
Use PPE (safety gloves, safety shoes, safety goggles).
Carry the unit with 2 persons (see the Technical Data for the
weight).
Transport the unit carefully on firm, level ground. The interior of
the unit can also be damaged by impacts or if it is dropped.
Do not loiter under the unit during transportation and operation.
The installation site should be a sufficiently large room to ensure
that it does not become too hot due to the heat emission.
The surface for the device should be flat and made of non-
flammable material.
A specific room size is prescribed for refrigerated circulators.
At high temperatures, position the unit under an exhaust hood as
much as possible due to potential vapors from the thermal bath
fluid.
Observe the safety sticker - do not remove!