Operation Of The Device And Resuscitation; Switching Primedic™ Heartsave Aed On And Off; Switching Primedic™ Heartsave Aed On; Switching Primedic™ Heartsave Aed Off - Primedic HeartSave AED Operating Manual

Hide thumbs Also See for HeartSave AED:
Table of Contents

Advertisement

Operating Manual
5
5.1
5.1.1
5.1.2
20439 / 03.04 / 03 / engl.

Operation Of The Device And Resuscitation

The process of resuscitation by this device has been devised in accordance with the
American Heart Association (AHA) and European Resuscitation Council (ERC) guide-
lines.
Switching PRIMEDIC™ HeartSave AED On And Off.
Switching PRIMEDIC™ HeartSave AED On
The device is activated automatically on removing the cover. If the device is not
switched on automatically, switch it on by pressing the
On / Standby button.
All buttons except the trigger button are then enabled.
Defibrillation is triggered only after ventricular fibrillation (VF) and ventricular tachy-
cardia (VT) have been detected.
Immediately after it has been switched, the device runs an internal self-test to check
the important functions and signal equipment. A signal tone confirms that the device
is ready for operation. Always check the functioning of the speaker.
Switching PRIMEDIC™ HeartSave AED Off
PRIMEDIC™ HeartSave AED can be switched off in many ways:
– By pressing the On / Standby button for about three seconds. At the same time,
there are three warning tones. This time is selected so that the device is not
switched off by mistake.
– By replacing the cover of the device.
– If the device does not detect a signal for ten minutes or if no button is pressed, it
is switched off automatically.
If the device detects any defect, it switches off automatically to prevent any possible
injuries.
HeartSave AED / AED-M
25

Advertisement

Table of Contents
loading

This manual is also suitable for:

Heartsave aed-m

Table of Contents