Safety Instructions For A Third Person; Safety Instructions For The Device - Primedic HeartSave AED Operating Manual

Hide thumbs Also See for HeartSave AED:
Table of Contents

Advertisement

Operating Manual
2.5
2.6
20439 / 03.04 / 03 / engl.
During defibrillation, detach all other electronic devices used for medical pur-
poses from the patient that do not have a defibrillation-proof user component.
Keep the defibrillation electrodes away from other electrodes and metallic parts
that are in contact with the patient!
Do not use the device for children under eight years or for children weighing less
than 25 kg!
Place the electrodes exactly as described
Before placing the defibrillation electrodes, dry the patient's chest and carefully
shave off any dense growth of hair.
Do not place the defibrillation electrodes directly over an implanted pacemaker.
This will cause the device to interpret incorrectly and damage the pacemaker due
to the defibrillation pulses.
Do not touch the patient during the ECG analysis and avoid any vibrations!
If the ECG analysis is conducted in a vehicle, then for a correct analysis, the ve-
hicle must be stopped and the engine must be switched off.
Stop resuscitation while PRIMEDIC™ HeartSave AED is carrying out the ECG
analysis.
Do not touch the patient during defibrillation! Avoid any contact between
– any parts of the patient's body (such as the bare skin of the head or the
legs), as well as
– conductive fluids (such as gels, blood or salt solution) and
– metallic objects in the patient's surroundings (such as the bed frame or
the stretcher) that represent an unintentional path for the defibrillation
current!

Safety Instructions For A Third Person

Before defibrillation, give a clear warning to any persons standing around to
move away from the patient and not touch him!

Safety Instructions For The Device

Any repairs, modifications, enhancements and installations of PRIMEDIC™
HeartSave AED must only be carried out by persons authorised and trained by
METRAX!
No parts of PRIMEDIC™ HeartSave AED can be repaired by the user!
The device must be equipped and operated with original accessories from
PRIMEDIC™ only!
Clean the device only when it has been switched off and the electrodes have
been removed and only in the instructed manner!
HeartSave AED / AED-M
7