Battery; Charging The Akupak - Primedic HeartSave AED Operating Manual

Hide thumbs Also See for HeartSave AED:
Table of Contents

Advertisement

4.4
4.5
20
HeartSave AED / AED-M
Change the power module only when the device is switched off and the connec-
tor of the defibrillation electrodes is removed.
Place the device on its rear side and press the release button (1) in the direction of
the arrow (1.) until the tab of the power module (2) is released and a part of the
power module (3) springs out of the compartment (4).
Turn the power module slightly in the direction of the arrow (2.) and pull it out of
the device in the direction of the arrow (3.).

Battery

The battery is a non-rechargeable Lithium battery and is fully charged on delivery.
This material is based on the latest technology and was selected due to its very long
durability and power storage function. The contacts are protected with a self-
adhesive sheet in order to prevent any automatic discharge or short circuit. Pull off
the protective sheet before inserting the battery.
Never try to recharge the battery. There is a risk of explosion.
Use the battery before the expiry date.
After using HeartSave, the battery should be replaced with a new battery. (to guaran-
tee a complete duration of operation on the next use)
Always refer to the supplement on the battery and keep it with the operating manual.
If the device has to be sent for servicing, remove the battery from the defibrillator
and put a scotch tape on the contacts.
Follow the special instructions for despatch while sending the battery.

Charging The AkuPak

The AkuPak can be charged in two ways:
– with the plug-in power pack and / or
– in the wall mount rack (optional).
The integrated end-of-discharge detection protects the accumulator from total dis-
charge. The end of charging is detected through the end-of-discharge voltage.
Charging the AkuPak at temperatures other than at the given operating temperatures
can damage the accumulator.
A completely empty accumulator must be charged for at least 30 minutes. Owing to
the type of the accumulator, a very small charging time can lead to errors in interpret-
ing charging by the device. The charging display can show a full accumulator by mis-
take. Safe operation of the device is also not guaranteed.
Operating Manual
20439 / 03.04 / 03 / engl.