Determining The Position Of Electrodes

Illustration 16
Position of Electrodes on the Patient
The electrodes are positioned
– in the region of the right breast, below the collarbone (1) and
– in the region of the left breast above the apex of the heart on the axillary line (2).

Opening SavePads

The tools holder contains one pair of electrodes, one shaver, one artificial respiration
cloth and one pair of disposable gloves.
Open the packing of the electrodes by tearing open the protective covering.

Removing the Chest Hair

If the patient has hair on the electrodes' positions, then it must be shaved off.
Shave off the hair at the electrodes' positions with the given shaver.