NanoDrop Lite Plus User Guide
Maintenance
28
NanoDrop Lite Plus User Guide
12. Repeat Steps
10
and
Tip To ensure accurate results, be sure to wipe the top and bottom of the
pedestal with a dry, lint-free laboratory wipe between measurements. Use a fresh
aliquot of PV-1 and a fresh pipette tip for each measurement.
After all 10 replicates have been measured, results will display as Pass/Fail. If the
performance verification results indicate Fail, thoroughly clean the pedestal following
the directions on the "Pedestal Cleaning and Reconditioning" document or the
Cleaning
and
Reconditioning
calibration check. If the calibration check results indicate Pass, proceed with
step
13.
13. Insert the USB memory device.
14. Select Export.
The following message will appear on the screen: Export in progress.....Please
do not remove USB drive until export is complete. When this message
disappears, it is safe to remove the USB memory device from the NanoDrop Lite
Plus.
After the data has been saved to the USB memory device, the data can be
transferred to a computer to archive or print. The performance verification is now
complete.
The verification report cannot be printed on labels using the optional NanoDrop Lite
Plus printer. Only new performance verification reports can be saved on USB
memory device and transferred to a computer for printing or archiving.
Only the most recent performance verification is saved to the instrument. Prior
performance verification history is overwritten each time a verification is performed.
The previous history cannot be retrieved.
To view the previous performance verificaiton
1. Go to the Home screen, select Maintenance.
11
until 10 measurements have been collected.
sections in the User Guide, and then repeat the entire