Thermo Scientific NanoDrop Lite Plus User Manual page 12

Micro-uv spectrophotometer
Table of Contents

Advertisement

1
NanoDrop Lite Plus User Guide
Nucleic Acid Measurements
12
NanoDrop Lite Plus User Guide
c
= (A *CF)/b
where,
c = the nucleic acid concentration in ng/μL
A = the sample absorbance
CF = the conversion factor in ng-cm/μL
b = the pathlength in cm
The generally accepted conversion factors for nucleic acids are:
• Double-stranded DNA: 50 ng-cm/μL
• Single-stranded DNA: 33 ng-cm/μL
• RNA: 40 ng-cm/μL
Table 3.Nucleic acid concentration range
Lower
Sample
Detection
Type
Limit
dsDNA
2.0 ng/μL
ssDNA
1.3 ng/μL
RNA
1.6 ng/μL
Note Absorbance reported is normalized to a 1.0 cm (10.0 mm) pathlength for
all measurements.
Upper
Typical reproducibility;
Detection
based on 10 replicates
Limit
(SD = ng/µL; CV=%)
1500 ng/μL
±2.0 ng/μL for sample
concentrations between 2.0 and
100 ng/μL samples; ±2% for
samples >100 ng/μL
990 ng/μL
±2.0 ng/μL for sample
concentrations between 2.0 and
100 ng/μL samples; ±2% for
samples >100 ng/μL
1200 ng/μL
±2.0 ng/μL for sample
concentrations between 2.0 and
100 ng/μL samples; ±2% for
samples >100 ng/μL
Thermo Scientific