Manufacturer's declaration – electromagnetic immunity
The PR 35 is designed to be operated under the electromagnetic conditions specified below. The customer
or the user of the device should ensure that it is used under these conditions.
Interference immunity
tests
Electrostatic discharge
(ESD), in accordance
with IEC 61000-4-2
Fast transient electric
bursts in accordance
with IEC 61000-4-4
Surges, in accordance
with IEC 6100-4-5
150
IEC 60601 – test level
± 6 kV Contact dischar-
ge
± 8 kV Air discharge
± 2 kV for power supply
cables
± 1 kV for input and
output cables
± 1 kV differential mode
voltage
± 2 kV common-mode
voltage
Compliance level
± 6 kV Contact dischar-
ge
± 8 kV Air discharge
± 1 kV for input and
output cables
Not usable
Electromagnetic conditions – guide-
lines
Floors should be made from wood
or concrete, or have ceramic tiles. If
the flooring is made from synthetic
material, the relative humidity must
be at least 30%.
The quality of the supply voltage
should correspond to that of a typical
business or hospital environment.
The quality of the supply voltage
should correspond to that of a typical
business or hospital environment.
Electromagnetic compatibility (EMC)