Paravan PR 35 User Manual page 151

Table of Contents

Advertisement

Voltage dips, short
interruptions and fluc-
tuations in the supply
voltage, in accordance
with IEC 61000-4-11
Magnetic field in the
power supply fre-
quency (50/60 Hz) in
accordance with IEC
61000-4-8
NOTE UT is the mains voltage before using the test level
Table 202, in accordance with DIN EN 60601-1-2, 6.8.3.201
Electromagnetic compatibility (EMC)
< 5% UT (>95% dip of
UT) for 0.5 periods
40% UT (60% dip of UT)
for 5 periods
70% UT (30% dip of UT)
for 25 periods
< 5% UT (>95% dip of
UT) for 5 seconds
3 A/m
Not usable
3 A/m
The quality of the supply voltage
should correspond to that of a typical
business or hospital environment.
If the movanimo user requires
continued functionality in the event
the power supply is interrupted, it is
recommended that the movanimo
be connected to an interruption-free
power supply or a battery.
Magnetic fields in the mains frequen-
cy should correspond to the typical
values which are present in a busi-
ness and hospital environment.
151