12.5 ELECTROMAGNETIC COMPATIBILITY

Despite all measures, interference and or EMC problems may occur.
Therefore, please note the following tables!
Further information on electromagnetic compatibility:
1. Medical devices are subject to precautionary measures in accordance with EMC and must
be installed and commissioned in accordance with the EMC information contained in the
operating instructions.
2. Portable and mobile RF communications equipment can affect electrical medical units.
3. The use of stands and mounting systems that are supplied by other manufacturers is not
permitted.
Emitted interference
GUIDELINES AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS
The MIMLED 600 and MIMLED 1000 minor surgical lights are intended for use in an
ELECTROMAGNETIC ENVIRONMENT as specified below. The user should ensure that the
units can be operated in such an environment.
Note: Home health care environments have higher immunity requirements compared to
professional healthcare facilities. Therefore, the requirements for professional health care
facilities regarding immunity to interference are included.
Phenomenon
Conducted and radiated
interference emissions
Harmonic distortion
Voltage fluctuations and
flicker
EMC
basic standard
IEC 61000-3-3
IEC 61000-3-2
IEC 61000-3-3
Immunity test level
Environment in areas of home health care
CISPR 11 Group 1, Class B
IEC 61000-3-2, Class A
47