Adjust Spring Force - MIMSAL MIMLED 600 Manual

Minor surgical lights
Hide thumbs Also See for MIMLED 600:
Table of Contents

Advertisement

10.2.5 ADJUST SPRING FORCE

Like any technical component, springs are subject to natural wear. Thus, the spring force may
decrease after longer operation and must be readjusted.
Adjust the spring force so that the spring arm with end unit remains in any desired position.
CAUTION: DESTRUCTION OF THE SPRING ARM
The spring force is adjusted in the upper end position.
ADJUST SPRING FORCE
CAUTION: DESTRUCTION OF THE SPRING ARM
Screwing in the adjusting screw too deeply will destroy the spring arm. Carefully
screw in the adjusting screw while repeatedly checking the braking force:
• If the spring arm lowers - the spring force is too low:
The adjusting screw must be turned to the left (counterclockwise).
• If the spring arm rises - the spring force is too high:
The adjusting screw must be turned to the right (clockwise).
Use the same procedure for the ceiling and wall mount!
30
1. Remove the joint cover on the left from the
direction of the end unit on the spring arm. To
do this, carefully lever the joint connection out of
the groove in the spring arm joint using a narrow
slotted screwdriver.
2. Move the end unit to the upper end position.
3. Insert slotted screwdriver into the hole.
4. Adjust spring force.
5. Affix the joint cover and snap it into place.

Advertisement

Table of Contents
loading

Table of Contents