MIMSAL MIMLED 600 Manual page 26

Minor surgical lights
Hide thumbs Also See for MIMLED 600:
Table of Contents

Advertisement

Use of circlip pliers with over-expansion
The illustration shows an example of circlip pliers with over-expansion guard (1).
The over-expansion guard prevents over-expansion of the circlip.
The circlip must not be dismantled without circlip pliers with over-expansion guard (1).
Follow the next steps:
1. Turn the adjusting ring (2) of the circlip pliers (1) until the circlip has an over-expansion guard
of approx. 8 mm as shown in the illustration.
2. This corresponds to an expansion of the circlip inner diameter of max. 32mm.
Figure. Use of circlip pliers with over-expansion guard
Disassemble/assemble circlip using circlip pliers with over-expansion guard
WARNING
An overexpanded circlip can cause the support arm system to crash:
• Carefully expand the circlip (1) only so far that it can just be guided over the
trunnion (2 ).
• To do this, expand the circlip (1) to a maximum inner diameter of 32.4mm. This
corresponds to an inner dimension of approx. 8mm between the eyelets.
1. Insert circlip pliers with over-expansion guard into the eyelets of the circlip (1).
2. Carefully expand the circlip (1) so that it can just be guided over the trunnion (2).
3. Carefully remove the circlip (1).
4. Mark circlip (1) for single use and keep for later assembly.
26