8. MAINTENANCE

Medical devices must be subjected to regular maintenance and testing cycles. This is
fundamental for compliance with safety concerns.
The manufacturer of the medical device is responsible for defining regular safety measures. The
operator is responsible for implementing these defined measures.
NOTE: Always switch the light to standby mode and disconnect the power plug or
disconnect the light from the power supply before carrying out any maintenance or
inspection work. Secure the light against unintentional restarting.
WARNING: RISK OF INJURY!
Support arm is spring-loaded and can spring up when the lamp head is removed.
When carrying out maintenance work, follow the manufacturer's instructions for
installation/dismantling (see chapter 12.6).

8.1 SUPPORT SYSTEMS

All support systems must be checked by the operator for the following points:
• Every six months:
1. Deformations of the support system
2. Cracks on plastic parts
3. Paint damage
• Yearly:
1. Extended check of the support system, e.g. holding force of the spring arm, check fastening
screw on underside of stand base and tighten if necessary.
2. Extended functional test such as ease of movement of the joints.
3. Electrical safety testing.
In the event of any malfunctions or damage, please notify your supplier.
WARNING: ELECTRIC SHOCK
Disconnect the unit from the power supply during all testing work.
19