Interference immunity
GUIDELINES AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS
The MIMLED 600 and MIMLED 1000 minor surgical lights are intended for use in an
ELECTROMAGNETIC ENVIRONMENT as specified below. The user should ensure that the
units can be operated in such an environment.
Note: The environment in home health care settings has higher immunity requirements
compared to professional healthcare settings. Therefore, the requirements for professional
health care facilities regarding immunity to interference are included.
Phenomenon
Static electricity discharge
Fast transient electrical
disturbances (Burst)
EN 61000-4-4
Surge voltage cable to
cable
Surge voltage cable to
earth
Voltage dips
Magnetic fields with
energy technology rated
frequencies
High-frequency
electromagnetic fields
Conducted interference
induced by high-frequency
Tabla. Emisiones de interferencia.
fields
Please note that the MIMLED 600 and MIMLED 1000 minor surgical lights will go out in
the event of a power failure or mains voltage drop. If uninterrupted illumination is required,
connect the lights to a power outlet with an emergency power function.
48
EMC
basic standard
IEC 61000-4-2
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-5
IEC61000-4-11
IEC 61000-4-8
IEC 61000-4-3
IEC 61000-4-6
Immunity test level
Environment in areas of home
health care
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±2 kV
100 kHz repetition frequency
±1 kV
± 2 kV
0% UT: 1 Period
70% UT: 25/30 periods
30 A/m
80 MHz – 2.7 GHz, 10 V/M
150 kHz – 80 MHz
10 Vrms