Operation; Check Before Each Use; Operation Of The Lamp Head - MIMSAL MIMLED 600 Manual

Minor surgical lights
Hide thumbs Also See for MIMLED 600:
Table of Contents

Advertisement

3. OPERATION

3.1 CHECK BEFORE EACH USE

1. Check the system for visible deformation. If such are detected, contact service immediately.
2. Ensure that the system has the required hygiene status for use.
3. Before each startup, the entire unit must be checked for proper functioning. The unit should
be moved in each degree of movement while checking the main function and the control
system.
4. If a lamp head is too difficult to move or no longer holds its position, the holding tensions can
be adjusted according to the operating instructions for supports and stands.
5. Check handle for cracks.
Do not operate the fixture if there is any doubt about its electrical safety or static and
dynamic stability.

3.2 OPERATION OF THE LAMP HEAD

The simple and ergonomic operating concept of the MIMLED enables intuitive operation by
the user.
The lamp head is connected to the support system via a bracket. The lamp head support allows
the light to rotate approximately 270° in its holder.
The 360° joint on the support system allows rotation around the horizontal axis.
The sterilizable handle allows both positioning of the lamp head and brightness adjustment of
the light during use.
Turning the sterilizable handle counterclockwise increases the illumination, while turning it
clockwise reduces it.
The blue "sterile handle" (the assembly consists of a sterilizable handle and clip) for the MIMLED
is not supplied sterile and must be disinfected and sterilized before first use.
The brightness indicators above the handle pulsate in standby mode with the lowest value
indication LED.
The light automatically goes into standby mode as soon as mains power is connected.
To completely switch off the light, the following must be taken into account,
depending on the type of installation:
Mobile stand: Pulling the mains plug switches the light off completely.
Wall mount: Pulling the wall power supply switches the light off completely.
Ceiling mount: Switching off the current-carrying socket by means of switch provided
by customer.
9