MIMSAL MIMLED 600 Manual page 51

Minor surgical lights
Hide thumbs Also See for MIMLED 600:
Table of Contents

Advertisement

The light system must be checked for the following points at the intervals specified below by
a company authorized by MIMSAL or by personnel with appropriate qualifications
Visual and functional check (annually checkout)
All parts are free from cracks*
The head is free from paint damage*
The function is proper*
Verification of electrical safety
All identification plates available/legible
Parts on the stand are deformation-free**
Mobility of the stand**
All joints operate smoothly**
Height stop of the stand correct**
Fuse segment of stand tested**
Stand locking ring in position**
Spring force set correctly**
Assessment for collision damage**
Tighten screw on stand base**
Electrical test protective conductor**
Electrical test leakage current**
The work is to be carried out including the necessary adjustment work and safety inspection.
*
Damaged or deformed components should be replaced as a precaution. Please contact the
supplier of the system.
Should one of the marked points be defective during the test, the system must be shut down
**
immediately as a precautionary measure in order to rule out further damage to people and
equipment. Inform the supplier of the system immediately.
The medical device book that belongs to every medical device and is prescribed by the
MPBetreibV must be kept on site. Service and maintenance work as well as safety checks must be
documented in this medical device book. Test reports such as this must be filed in the respective
medical device book.