11.1 CHOICE OF FASTENERS

NOTE
1. The person responsible for the installation is responsible for the safe selection of
the fasteners and the safe execution of the fastening.
2. For lightweight walls, we recommend fastening with a counter-plate (not included
in delivery).
WARNING: LOAD DATA
• The load torque on the spring arm of the wall mount must not exceed 39 Nm.
• The load torque on the spring arm of the ceiling mount must not exceed 30 Nm.
• No safety factors are included in the specified load data. The prescribed regional
safety factors shall be included.
• The load data of the wall and ceiling unit can be taken from the following table.
LOAD DATA FOR WALL MOUNT
Load torque wall mount
Vertical weight force
Pull-out force per wall plug (total 2 pcs.)
LOAD DATA FOR CEILING MOUNT
Load torque ceiling mount
Vertical weight force
Pull-out force per wall plug (total 4 pcs.)
Table. Load data for wall and ceiling mount

11.2 WALL MOUNTING

NOTE: A properly grounded socket in the area of the connection line is required for
the mains connection of the wall mount.
ACCESSORIES
Washer
Tab washers
Circlip held on collet
1x 4520.12r008/00 Switching power supply
1x 4520.12-006/00 Bend protection sleeve White
1x 4520.12-004/00 Strain relief clamp 14.42.770
2x 4520.20-007/00 Connection terminal two-pole with lever
2x 0606.01-004/00 Hexagon nut DIN EN 24035 M4-04
1x 4510.20-014/00 Folding ferrite WE 74271131
110 Nm
97 N
624 N
85 Nm
147 N