8.2 LAMP HEAD

The following inspections/maintenance must be performed annually:
1. Checking for cracks, deformations in plastic parts and seals.
2. Electrical safety testing.
3. Extended function test.
4. Paint damage.

9. DISPOSAL

The light may be operated for a maximum of 10 years after initial commissioning. The unit
should then be taken out of service and disinfected and cleaned for disposal. Therefore, for
proper disposal, contact an approved disposal company. For proper disposal of the system,
please contact an authorized disposal company.
NOTE: Do NOT disassemble the spring arms and articulated joints. Some of the
spring arms and joints contain pretensioned compression springs which can release
their tension abruptly if they are not dismantled properly.
NOTE: Do not dispose of the product with normal domestic waste.
Perform all disinfection or sterilization measures prior to decommissioning to exclude
contamination of the environment.