In the event of service or maintenance, a new, unused circlip must always be fitted.
Insert circlip pliers with over-expansion guard into the eyelets of the circlip (1).
The circlip must lie fully and straight in the groove provided. This is checked by the following
steps: The circlip (1) must audibly engage in the groove on the trunnion.
Figure. Mount circlip
Acoustic test
• The circlip (1) must audibly engage in the
groove on the trunnion.
Visual inspection
• The washer Ø39 mm (2 ) must be mounted
under the circlip (1).
• The circlip (1) must not be non-circular.
• The distance between the two eyelets
in the circlip (1) must correspond to the
distance in the unstressed state. A higher
distance indicates that the circlipis not
mounted correctly.
Rotating the circlip in its groove
• Place a small, suitable screwdriver on the
eyelet (1) of the circlip (2 ) and carefully
turn the circlip (2 ) in the direction of the
arrow.
• Be careful not to widen the circlip (2 ) or
push it out of the groove.