Condition Errors 301 To 399, Uncalibrated Integrity - Keysight N9000A Service Manual

Cxa signal analyzer
Hide thumbs Also See for N9000A:
Table of Contents

Advertisement

Instrument Messages

Condition Errors 301 to 399, Uncalibrated Integrity

Condition Errors 301 to 399, Uncalibrated Integrity
Errors with measurement calibration/alignment routines and signals.
An event with the error number shown in the table means the condition has been
detected.
When the condition is cleared, an event with the error number plus 1000 is
generated.
For example, error 307 indicates an "AC: unspec'd below 10 MHz" condition has
been detected; error 1307 indicates that failure has been cleared.
An E in the Error or Warning column means that an error is put up on the front
panel and sent out to SCPI when this condition is detected. A W in this column
means that a Warning is put up on the front panel but nothing goes out to SCPI.
Nothing in this column means nothing is put up; status bit only.
Err#
Message
301
Meas Uncal
303
Sig ID On
305
No Long Code Phase
307
AC coupled: Accy
unspec'd <10 MHz
309
User cal
309
User Cal; Adjusted for
new RBW
116
Error or
Verbose/Correction Information
Warning
W
W
W
W
AC input coupling will function at lower frequencies, but
the performance is not specified below 10 MHz.
W
W
The measurement RBW has been changed since the last
calibration (~CAL)
Chapter 3

Advertisement

Table of Contents
loading

Table of Contents