Calibration for NA, CAT, and VVM Modes

Verifying Calibration and Jumper Cable Integrity

Verifying Calibration and Jumper Cable Integrity
After calibrating, it is important to verify that the calibration is good. When
using a jumper cable, also verify that the cable is of high quality.

Verify a Calibration

— Connect a LOAD standard at the calibration reference plane (where
— In CAT Mode, select
— In NA Mode, select
— For frequencies lower than 9 GHz, observe the trace on the FieldFox
Verifying Phase Accuracy
— Connect a LOAD standard at the calibration reference plane (where
— In NA Mode, select a S11 Reflection with Polar or Smith Chart format.
— Because all LOAD standards have delay, you should see a small amount of

Test the Jumper Cable

With the LOAD standard still connected, move the jumper cable while
observing the trace.
— If the measurement trace is relatively stable, the jumper cable is of good
— If you observe significant movement in the peaks of the measurement trace
142
calibration standards were connected). This may be at the end of the
jumper cable or at FieldFox PORT 1 connector.
Return
S11
screen:
— Lower than 35 dB indicates a GOOD calibration.
— Higher than 30 dB indicates that you should recalibrate.
calibration standards were connected).
phase rotation as a function of frequency. In general, the measurement
result should agree with the characteristics of the calibration standard.
quality.
when moving the cable (>5 dB), the jumper cable may need to be replaced.
Loss.
with Log Mag format.
Keysight N9927-90001 User's Guide