Amplitude Accuracy At 50 Mhz - Keysight N9000A Service Manual

Cxa signal analyzer
Hide thumbs Also See for N9000A:
Table of Contents

Advertisement

Functional Tests

Amplitude Accuracy at 50 MHz

Amplitude Accuracy at 50 MHz
Test Limits
Amplitude Accuracy should remain within 1.13 dB of the measured source value
across the range of source levels and changes in resolution bandwidth. The Preamp
(option P03, P07) should remain within ±1.3 dB of measured values.
A synthesized sweeper is used as the signal source for the test. The source
amplitude is varied using the signal source amplitude control. The attenuation and
resolution bandwidth are varied on the signal analyzer. The source amplitude is
measured by the power meter and signal analyzer at each setting, and the values
compared. The difference between each pair of measurements indicates the
amplitude accuracy.
Item
Adapter
Type-N(m), to 3.5 mm(m)
Adapter
3.5 mm(f) to 3.5 mm(f)
Attenuator, 30 dB
Accuracy: < 0.5 dB at 50 MHz
Cable
3.5 mm, 1 meter
Cable
BNC, 120 cm
Power Meter
Compatible with power sensor
Amplitude Range: −25 dBm to 10 dBm
Power Sensor
Power Splitter
3.5 mm (f) connectors
6 dB loss
Synthesized Sweeper
Typical Temperature Stability: 0.01 dBc/°C
300
Critical Specifications
Recommended
Agilent Model
1250-1743
83059B
11708A
11500E
10503A
E4418B
8485A
11667B
PSG
Chapter 15