Download Print this page

Emerson Rosemount Wireless ET410 Quick Start Manual page 21

Corrosion transmitter
Hide thumbs Also See for Rosemount Wireless ET410:

Advertisement

April 2023
9. If there is an excessive length of spare strap, you may cut off
the excess length.
10. Click the Complete button.
Verify the Install State is Off and Installed is selected in the
footer of the application.
Figure 3-6: Installation Tool Screen: Fully Provisioned
11. Remove the CC21 and fit the power module, tightening the
two power module retaining bolts. Refer to the
BP20E Power Module for Wireless Corrosion Transmitter Quick
Start
When the power module is fitted, the sensor automatically
restarts and tries to join the WirelessHART
network of 100 sensors this can often take two hours, and
sometimes up to six hours.
3.5 
Fitting the lanyard
 WARNING
Use supplied lanyard to prevent sensor falling from heights,
potentially causing injury.
Procedure
1. Wrap the lanyard around the circumference of the pipe, on top
of any cladding.
Note
The 6.6 ft (2m) lanyard is sufficient for a pipe diameter up to
20 inches. When it is not possible to wrap the lanyard around a
pipe, find an alternative attachment point for the lanyard.
Tip
For pipes exceeding 20 inches in diameter, lanyards may be
linked together. For sensors in close proximity to each other, a
single lanyard may be used.
Quick Start Guide
If it is poor, slacken off the strap and adjust the sensor
position before returning to Step 3.
If the waveform quality is good, continue to Step 9.