Operator's Manual
Performance Verification
This procedure can be used to verify the warranted characteristics of the probe. The
recommended calibration interval for the models CP030, CP030-3M, CP030A, CP031,
and CP031A Current Probes is one year. The complete performance verification
procedure should be performed as the first step of annual calibration. Performance
verification can be completed without removing the probe covers or exposing the user
to hazardous voltages. Test results can be recorded on a photocopy of the Test Record
provided at the end of the manual.
The warranted characteristics of the probe are valid at any temperature within the
Environmental Characteristics listed in the Specifications. However, some of the other
test equipment used to verify the performance may have environmental limitations
required to meet the accuracy needed for the procedure. Make sure that the ambient
conditions meet the requirements of all the test instruments used in his procedure.
NOTE: Operation of the probe as described requires software version 4.3.0.0 or higher
for CP03x and 7.8.x.x or higher for CP03xA. To confirm the version installed, choose
Utilities > Utilities Setup from the oscilloscope menu bar, then open the Status tab.