Manual Control
Focusing
Turning counterclockwise = downwards
Turning clockwise = upwards
Toggle between
coarse and fine focusing
1. Press the black button briefly.
When the system is switched on, the
coarse focus is always switched on also.
Leica Z6 APO A / Z16 APO A
Information about the focus positions
Focus positions can be stored using the
★
manual control or on the computer.
A total of up to 5 focus positions can be
★
stored.
If a sixth focus position is stored, the one
★
nearest to it will be deleted.
Each focus position can be canceled indivi-
★
dually.
The system will travel to each of the indi-
★
vidual focus positions in the sequence in
which they were stored.
Switching off the system cancels all of the
★
stored focus positions.
User Manual
Saving the focus positions
1. Focus on the first specimen detail.
2. Press the red switch for at least 1.5 seconds.
Short beeps provide confirmation.
3. Focus on the next specimen details and
save these positions also.
Moving to stored focus positions
1. Press the red switch briefly.
Deleting stored positions
Press the red switch until you hear the audible
signal: 2 short beeps – pause – 2 long beeps.
84