Leica IsoPro™ Motorized Mechanical Stage: Assembly (cont'd.)
The bases
The TL BFDF, TL RC™ and RCI™ transmitted-light
bases can be equipped with three different
stages: Standard stage, manual and automated
IsoPro™ mechanical stage. The selected stage is
mounted on the base before commissioning.
You can switch between the stages at any time
with just a few hand movements.
The following paragraph assumes use of the
base without the stage mounted. Disassembly
is performed in reverse order of the following
steps.
Leica Z6 APO A / Z16 APO A
Mechanical stage and base
 
Never move the sledge of the motorized
mechanical stage manually in the X
direction, as otherwise the mechanical system
will be damaged!
1. Place the mechanical stage onto the base.
2. Pull the upper part of the mechanical stage
carefully towards the user and fasten the
lower part onto the transmitted-light base.
3. Attach the mechanical stage evenly to the
three threaded holes.
4. Now move the mechanical stage as far as it
will go in the direction of the column.
5. Insert the glass plate into the mechanical
stage.