Thermo Scientific EASYpure 1164 Series Operation Manual And Parts List page 25

Table of Contents

Advertisement

Warning
Depressurize system prior to opening
cartridge access door.
Caution
Do not unscrew the cover, as this will
loosen the water tight seal and may
damage the replacement lamp. Pull the
cover straight up.
Note
Pay close attention to the orientation
of the plug as it is attached to the UV
lamp. It will be necessary to ensure
that the plug is installed in the same
manner as removed.
M
3.
Open the cartridge access door. Remove the
screws securing the EASYpure UV/UF cover.
4.
Remove the cover by lifting straight up.
5.
Locate the UV oxidation chamber and pull the
top black plastic cover off. Do not pull on the
cable.
6.
While holding on to the lamp, remove the plug
from the lamp. Dispose of the UV lamp appro-
priately. The used lamp does contain mercury
vapor and should not be disposed of in the
trash. Recycle or dispose of the used lamp as
hazardous waste.
7.
Remove the replacement lamp from its contain-
er. DO NOT TOUCH THE GLASS PORTION
OF THE LAMP. It is recommended that lint free
gloves be worn when handling the lamp. The
glass portion must be free of fingerprints, per-
spiration, etc. Even a light coating of perspira-
tion will reduce the effectiveness of the lamp.
8.
Clean the lamp with isopropyl alcohol and a lint
free cloth.
9.
Carefully insert and hold the UV lamp partially
into the UV chamber.
10. Connect the UV lamp to the receptacle in the
black plastic cover of the UV chamber. Replace
the black plastic cover on the UV chamber.
11. Reinstall the EASYpure UV/UF cover. Close the
cartridge access door.
12. Reattach the power cord and the pump protec-
tor or the pump interlock and reconnect the unit
to the power supply.
13. Operate normally.