Shutdown - Thermo Scientific EASYpure 1164 Series Operation Manual And Parts List

Table of Contents

Advertisement

16. Reinstall the screw holding the cell-cable retain-
ing clip. Reinstall membrane switch lead. Re-
place the front cover. Replace the Luer fitting
with new Teflon ® tape and reinstall fitting.
17. Reinstall the EASYpure UV/UF cover. Close the
cartridge access door.
18. Reattach the power cord and the pump protector
or the pump interlock and reconnect the unit to
the power supply.
19
Operate normally.

Shutdown

If the EASYpure UV/UF is to be shut down for an extended
period of time, the unit should be completely drained and
the cartridges and ultrafilter removed to prevent the growth
of bacteria.
If the system has remained inactive and full of water for
more than 96 hours, the unit should be drained, sanitized
and new cartridges installed prior to use.