Figure 7: Orientation of Swivel Base
Pump Interlock
Plug
Main Power
Switch
Power Entry
Module
Figure 8: EASYpure UV/UF Connections
Door Latch
Feedwater Inlet
Connection
Reject Water
Connection
Reject Water Connection
When the EASYpure UV/UF flushes its membrane, the
water used is sent to drain through this connection. To
install:
1.
Locate the reject water tubing. This is the
piece of 1/4" O.D. tubing that has an adapter
on one end and no connector on the other
end. Remove the nut, grab ring, backup ring
and o-ring from the reject water fitting located
at the lower right of the unit. (See Figure 3 for
part identification and Figure 8 for fitting loca-
tion.)
2.
Follow instructions under Tubing Adapter In-
stallation and connect the reject water tubing
to the reject connection on the EASYpure
UV/UF.
3.
Route the other end of the reject water tubing
(the end with the adapter installed) to an at-
mospherically vented drain. Ensure there are
no kinks in the tubing and that it proceeds in
a downward plane. If possible, use the tubing
adapter to permanently install the reject water
tubing in the drain.