Thermo Scientific EASYpure 1164 Series Operation Manual And Parts List page 19

Table of Contents

Advertisement

High Purity/Low TOC Cartridge
D50229
Figure 10: Proper Cartridge Position and Order
Warning
Use properly grounded electrical outlet
of correct voltage and current handling
capacity.
Caution
Do not allow EASYpure UV/UF to oper-
ate unless water is available to unit.
Note
For more demanding applications
where low TOC water is required, a
rinse of 15-20 liters through the car-
tridges and filter may be necessary.
Pretreatment Cartridge
D50230
ULTRApure Cartridge
D50233
10. Press the "Flush" button twice to begin an ex-
tended flush. When water begins exiting the
system, close the draw-off valve.
11. When the extended flush ends, press the
START/STOP button to stop and depressur-
ize the system by opening the draw-off
valve and allowing water to drain from the
unit until draining ceases. Close the draw-
off valve. Open the cartridge access door and
repeat steps 2 - 6 with the EASYpure High
Purity/Low TOC (Catalog No. D50229) car-
tridge, placing it in the left-hand position. Do
not install the 0.2 micron filter and bell assem-
bly at this time.
12. Close the cartridge access door. Press the
"Flush" button twice again to begin another
extended flush. When the extended flush
ends, open the draw-off valve and press the
START/STOP button. Rinse 10-15 liters of
water through the draw-off valve to drain.
Close the draw-off valve.
13. Remove a new 0.2 micron filter and bell as-
sembly from its bag and insert it into the Luer
fitting on the draw-off valve. Gently turn it
clockwise until it is fully seated in the Luer fit-
ting.
14. Remove the protective cap from the filter bell.
Open the draw-off valve and flush 10-15 liters
of water through the 0.2 micron filter.
I
O
NITIAL
PERATION
19