Dräger Polaris 600 Instructions For Use Manual page 69

Surgical light
Hide thumbs Also See for Polaris 600:
Table of Contents

Advertisement

No.
LED
25
LED
LEDs on the mains connection
component (signal combinations)
No.
LED
Mains power
20
supply (Sup-
ply mains)
Emergency
power supply
21
(Emergency
power)
Output volt-
age to light
22
(Output volt-
age)
Notify Dräger-
24
Service (Con-
tact service)
Instructions for use Polaris 600
Status
Off
Pulsating yellow
Status
Flashing yellow
Flashing yellow
Flashing yellow
Continuously yellow
Cause
Processor in the mains connection
component is faulty
Processor in the mains connection
component is fault-free
Cause
Internal fault in the mains connection
component
Input source is unknown
Output voltage level is unknown
Troubleshooting
Remedy
Inform Dräger-
Service
No remedy nec-
essary
Remedy
Notify Dräger-
Service
69