Dräger Polaris 600 Instructions For Use Manual page 76

Surgical light
Hide thumbs Also See for Polaris 600:
Table of Contents

Advertisement

Maintenance
Component
Complete light and arm
system
76
Interval
Every 2 years
Task
Check the locking seg-
ments on the swivel arm,
see "Checking the locking
devices on the C swivel
arm and the S swivel arm"
on page 86.
Check the locking seg-
ments on the spring arm,
see "Checking the locking
devices on the spring arm"
on page 87.
Check the circlip on the
lower swivel arm to the
spindle joint, see "Check-
ing the circlip on the lower
swivel arm to the spindle
joint" on page 87.
Grease the brake screws,
see "Greasing the brake
screws" on page 87.
– Grease the brake
screws on the swivel
arm and the spring
arm.
Check the load balancing
and spring force on the
spring arm and readjust if
necessary.
Check the arm system
(central axis, swivel arm,
and spring arm) for colli-
sion damage (welding
seams free of cracks).
Check for deformations of
the arm system and
whether parts have
become loose.
Instructions for use Polaris 600
Personnel responsible
Specialized service per-
sonnel